人妻中出视频

Skip to main content

Shock to the System: Certain Wearable Gadgets Could Interfere With Implantable Cardiac Electronic Devices

Benjamin Sanchez Terrones and Benjamin Steinberg
Research by Benjamin Sanchez Terrones, Ph.D., and Benjamin Steinberg, M.D., suggests that some wearable smart devices could interfere with cardiac implantable electronic devices such as pacemakers and implantable cardioverter defibrillators.

In this high-tech era, wearable devices such as smartwatches have proven to be invaluable companions for the health conscious. But a new study from the 人妻中出视频 of Utah shows that for a small group of people, some of these electronic fitness gadgets could possibly be risky to their health 鈥 even potentially deadly.

人妻中出视频 of Utah electrical and computer engineering assistant professor  and U of U 人妻中出视频 associate professor of medicine Benjamin Steinberg, M.D., have published a new study that shows wearable devices such as the Samsung Galaxy watch 4, Fitbit smart scales, or Moodmetric smart rings, among others, have sensing technology that could interfere with cardiac implantable electronic  devices (CIEDs) such as pacemakers, implantable cardioverter defibrillators (ICDs), and cardiac resynchronization therapy (CRT) devices.

鈥淭his study raises a red flag,鈥 says Sanchez Terrones. 鈥淲e have done this work in simulations and benchtop testing following Food and Drug Administration accepted guidelines, and these gadgets interfere with the correct functioning of the CIEDs we tested. These results call for future clinical studies evaluating the translation of our findings to patients wearing CIEDs and using these wearable devices.鈥

Their study was published in the newest edition of the scientific journal, Heart Rhythm. The paper, authored by U electrical and computer engineering graduate student Gia-Bao Ha, Sanchez Terrones, Steinberg, U internal medicine professor Roger Freedman, and Universitat Aut貌noma de Barcelona cardiology professor Antoni Bay茅s-Gen铆s, can be read here.

At issue are specific wearable smartwatches, at-home smart scales, and smart rings that utilize bioimpedance, a type of sensing technology that emits a very small, imperceptible current of electricity (measured in microamps) into the body. For smartwatches like the Samsung Galaxy Watch 4 or the Fitbit Aria 2 smart scale, the electrical current flows through the body and the response is measured by the sensor to determine the person鈥檚 body composition such as skeletal muscle mass or fat mass. For smart rings like the Moodmetric smart ring, bioimpedance sensing technology is used to measure a person鈥檚 level of stress.

But after conducting comprehensive testing of bioimpedance on three cardiac CRT devices from manufacturers Medtronic, Boston Scientific, and Abbott, Sanchez Terrones鈥 team learned the slight electrical currents from these wearable gadgets can interfere and sometimes confuse cardiac implantable devices into operating incorrectly.

In the case of a pacemaker, which sends small electrical impulses to the heart when it is beating too slowly, the bioimpedance鈥檚 tiny electrical current could trick the heart into thinking it is beating fast enough, preventing the pacemaker from doing its job when it is supposed to.

鈥淲e have patients who depend on pacemakers to live,鈥 Steinberg, a cardiac electrophysiologist, says. 鈥淚f the pacemaker gets confused by interference, it could stop working during the duration that it is confused. If that interference is for a prolonged time, the patient could pass out or worse.鈥

For other types of medical devices such as implantable cardioverter-defibrillators, which not only act as a pacemaker but can also shock the heart to restore a regular heart rhythm, a wearable device with bioimpedance could trick the defibrillator into delivering the patient an unneeded electric shock, which can be painful.

Nearly all if not all implantable cardiac devices already warn patients about the potential for interference with a variety of electronics due to magnetic fields, such as carrying a mobile phone in your breast pocket near a pacemaker. But Sanchez-Terrones says this is the first time a study has discovered problems associated with a gadget鈥檚 bioimpedance sensing technology.

鈥淭he scientific community doesn鈥檛 know about this,鈥 he says. 鈥淣o one has looked at whether this is a real concern or not.鈥

Sanchez Terrones and Steinberg emphasize that the research does not convey an immediate or clear risk to patients who use these kinds of wearable devices, but they believe it鈥檚 a first step for further study.

鈥淲e need to test across a broader cohort of devices and possibly in patients with these devices,鈥 Steinberg said.

鈥淯ltimately, more studies are needed to evaluate the clinical translation of our findings and ensure the health of our patients,鈥 Sanchez Terrones added.